COMMAND REFERENCE

4 - 31

ABORt

Syntax: ABORt

Description:

The ABORt command resets the trigger system and places it in the "IDLE" state. Pending actions that were already started are finished immediately. The ABORt command is not finished until the pending actions have been terminated.

Note: The commands *RST and ABORt have the same effect on the trigger functions, except that ABORt does not affect the state of the INITiate:CONTinuous command. So, when an ABORt command is sent while the INITiate:CONTinuous is ON, the trigger system will leave the IDLE state at once.

Example:

 

Send ABORt

Aborts the current acquisition.

Send CONFigure:AC

Configures for AC-RMS value.

Send READ:AC?

Initiates and reads the AC-RMS value.

Read <the measured AC-RMS value>

Page 131
Image 131
Fluke PM-3390B, PM-3380B Syntax ABORt Description, Aborts the current acquisition, Configures for AC-RMS value