COMMAND REFERENCE

4 - 117

TRIGger[:SEQuence[1]]:FILTer:LPASs:FREQuency

TRIGger[:STARt]:FILTer:LPASs:FREQuency

TRIGger[:SEQuence[1]]:FILTer:LPASs:STATe

TRIGger[:STARt]:FILTer:LPASs:STATe

Syntax:

TRIGger[:SEQuence[1]]:FILTer:LPASs:FREQuency <NRf>

 

 

 

 

MINimum MAXimum

 

TRIGger[:SEQuence[1]]:FILTer:LPASs:STATe <Boolean>

Alias:

TRIGger[:STARt]:FILTer:LPASs:FREQuency <NRf>

 

 

 

 

MINimum MAXimum

 

TRIGger[:STARt]:FILTer:LPASs:STATe <Boolean>

 

<NRf>

The cutoff

frequency expressed in hertz. Possible

 

 

values are:

 

 

-

0

Defines trigger DC coupling (MINimum).

 

 

-

10

Defines trigger AC coupling.

 

 

- 30000 Defines LF-reject (MAXimum).

 

<Boolean>

0 OFF

Sets low-pass filter off.

 

 

1 ON

Sets low-pass filter on.

Query form: TRIGger[:SEQuence[1]]:FILTer:LPASs:FREQuency?

 

 

 

 

MINimum MAXimum

Alias:

TRIGger[:STARt]:FILTer:LPASs:FREQuency?

 

 

 

 

MINimum MAXimum

Response:

0.00E+00 1.00E+01 3.0E+04

 

If MINimum was specified, the minimum possible cutoff frequency

is returned, i.e., 0 Hz. If MAXimum was specified, the maximum possible cutoff frequency is returned, i.e., 30 KHz.

Query form: TRIGger[:SEQuence[1]]:FILTer:LPASs:STATe?

Alias: TRIGger[:SEQuence[1]]:FILTer:LPASs:STATe?

Response: 0 1

0High-pass filter active (HF-reject).

1Low-pass filter active.

Page 217
Image 217
Fluke PM-3370B TRIGgerSEQuence1FILTerLPASsFREQuency NRf, TRIGgerSEQuence1FILTerLPASsSTATe Boolean, 00E+00 1.00E+01 3.0E+04