COMMAND REFERENCE

4 - 65

INPut<n>:POLarity

Syntax: INPut<n>:POLarity NORMal INVerted

<n> 2 4

Note: Input 4 is not applicable for PM33x0B.

Query form: INPut<n>:POLarity?

<n> 2 4

Response: NORM INV

Description:

The INPut<n>:POLarity command sets the polarity of the signal on the input channels two and four. The polarity can be set to NORMal (default) or INVerted (inverted signal).

After a *RST command, the polarity is NORMal.

Example:

 

Send *RST

Resets the instrument.

Send CONFigure:AC (@2)

Configures channel 2.

SendÆ SENSe:FUNCtion "XTIMe:VOLTage2"

Sets channel 2 ON.

Send INPut2:COUPling DC

Sets DC input coupling on.

Send INPut2:POLarity INVerted

Sets INV CH2 on.

Send READ:DC? (@2)

Requests DC channel 2.

Read <DC value>

Reads inverted DC value.

Front panel compliance:

The INPut<n>:POLarity command is the remote equivalent of the front panel INV keys.

Page 165
Image 165
Fluke PM-3380B, PM-3390B, PM-3370B, PM-3394B, PM-3384B user manual INPutnPOLarity, Response Norm INV Description