COMMAND REFERENCE

4 - 109

TRACe:COPY

Syntax: TRACe:COPY <destination_trace>,<source_trace>

Alias: DATA:COPY <destination_trace>,<source_trace>

<source_trace>

CHn Mi_n

<destination_trace> Mi_n

n = 1 .. 4

i = 1 .. 8 (standard memory)

i = 1 .. 50 (extended memory)

Description:

Copies a trace from one trace memory (source) to another (destination). The contents of the <source_trace> is copied to the <destination_trace>. The trace administration data is copied as well. If the oscilloscope is in the analog mode, error -221 "Settings conflict;Digital mode required" is generated.

Note: If the <source> trace is being built, the copy action takes place after completion of the source trace.

Limitations:

For the PM33x0B CombiScope instruments:

-CH3 and Mi_3 is not applicable.

-CH4 is the external trigger view channel, so:

EXTernal is the alias for CH4.

Mi_E is the alias for Mi_4.

Example:

Send *RST

Send SENSe:FUNCtion "XTIMe:VOLTage2"

Send TRACe:COPY M1_1,CH1

Front panel compliance:

Resets the instrument. Switches channel 2 on. The result is that the trace memories of channel 1 and 2 copied to M1_1 and M1_2 respectively.

The TRACe:COPY command is the remote equivalent of the front panel COPY option of the SAVE menu.

Page 209
Image 209
Fluke PM-3384B, PM-3380B, PM-3390B, PM-3370B, PM-3394B user manual TRACeCOPY, Limitations