4 - 64

COMMAND REFERENCE

INPut<n>:IMPedance

Syntax: INPut<n>:IMPedance <NRf> MINimum MAXimum

<n>

[1]

2 3 4

<NRf>

50

1E6

<MINimum>

Equals 5.00E+01 (50Ω)

<MAXimum> Equals 1.00E+06 (1 MΩ)

Note: Channel 3 is not applicable for PM33x0B.

Query form: INPut<n>:IMPedance? [MINimum] [MAXimum]

<n>

[1] 2 3 4

Response: 5.00E+01 1.00E+06

If <MINimum> was specified, 5.00E+01 (50Ω) is returned.

If <MAXimum> was specified, 1.00E+06 (1 MΩ) is returned.

Description:

The input impedance of channel <n> is set according to the impedance parameter specified. The impedance can be specified low (50Ω) or high (1 MΩ).

After a *RST command, the impedance is 1 MΩ.

Restrictions:

The impedance of the following input channels is fixed at 1 MΩ:

-All channels of the PM3384B CombiScopes instruments.

-Channel 4 of the PM33x0B CombiScope instruments.

Example:

Send INPut2:IMPedance 50 Selects 50Ω input impedance for channel 2.

Front panel compliance:

The INPut<n>:IMPedance command is the remote equivalent of the front panel 50Ω option in the VERT MENU menu.