3 - 66

USING THE COMBISCOPE INSTRUMENTS

3.12 Print/Plot Functions

The HCOPy:DEVice <TYPE> command allows you to select a hardcopy device. The following selections can be made:

DEVICE:

TYPE:

NOTE:

Plotter

HPGL

HPGL plot data format

Plotter

HP7440

 

Plotter

HP7550

 

Plotter

HP7475A

 

Plotter

HP7470A

 

Plotter

PM8277

 

Plotter

PM8278

 

Printer

FX80

Epson FX80 compatibles (9 points)

Printer

HP2225

ThinkJet

Printer

LQ1500

Epson LQ150 compatibles (24 points)

Printer

HPLASER

HP LaserJet series II & III

Printer

HP540

HP DeskJet (new style protocol)

Generator

DUMP_M1

Trace dump to one of the arbitrary waveform

 

 

generators PM5138, PM5139, or PM5150.

The HCOPy:DATA? query allows you to request a hardcopy of the picture on the screen of your CombiScope instrument. The response data is formatted according to the current printer/plotter options, which can be selected via the front panel UTILITY menu. After a *RST command, the option "plotter; HPGL" is selected.

The response data to a HCOPy:DATA? query can be sent to a connected plotter or printer to make a hardcopy. The response data is sent as block data of indefinite length and is therefore, preceded by the preamble #0 of 2 bytes. This preamble must be removed from the beginning of the block data, before sending it to a plotter or printer device.

Page 85
Image 85
Fluke PM-3380B, PM-3390B, PM-3370B, PM-3394B, PM-3384B user manual Print/Plot Functions, Device Type, Hpgl, Hplaser, DUMPM1