Fluke PM-3380B, PM-3390B Pass Boolean, Sets high-pass filter off, Sets high-pass filter on

Models: PM-3394B PM-3390B PM-3370B PM-3380B PM-3384B

1 297
Download 297 pages 56.64 Kb
Page 215
Image 215

COMMAND REFERENCE

4 - 115

TRIGger[:SEQuence[1]]:FILTer:HPASs:FREQuency

TRIGger[:STARt]:FILTer:HPASs:FREQuency

TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe

TRIGger[:STARt]:FILTer:HPASs:STATe

Syntax: TRIGger[:SEQuence[1]]:FILTer:HPASs:FREQuency <NRf>

MINimum MAXimum

TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe <Boolean>

Alias: TRIGger[:STARt]:FILTer:HPASs:FREQuency <NRf>

MINimum MAXimum

TRIGger[:STARt]:FILTer:HPASs:STATe <Boolean>

<NRf>

The cutoff frequency expressed in hertz. The only

 

possible value is 30000, which defines HF-reject (LF-

 

pass).

 

<Boolean>

0

OFF

Sets high-pass filter off.

 

1

ON

Sets high-pass filter on.

Query form: TRIGger[:SEQuence[1]]:FILTer:HPASs:FREQuency?

MINimum MAXimum

Alias: TRIGger[:STARt]:FILTer:HPASs:FREQuency?

MINimum MAXimum

Response: 3.00E+04 1.00E+08 2.00E+08

3.00E+04 Fixed cutoff frequency of 30 KHz (MINimum).

1.00E+08 Bandwidth of 100 MHz (MAXimum for PM338xB).

2.00E+08 Bandwidth of 200 MHz (MAXimum for PM339xB).

Query form: TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe?

Alias: TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe?

Response: 0 1

0Low-pass filter active.

1High-pass filter active (HF-reject).

Page 215