3 - 46

USING THE COMBISCOPE INSTRUMENTS

SENSe

TRACe

 

 

 

 

CALCulate

CH 1

M1_1

M2_1

M3_1

M50_1

CALCulate[1]

CH 2

M1_2

M2_2

M3_2

M50_2

 

CH 3

M1_3

M2_3

M3_3

M50_3

 

CH 4

M1_4

M2_4

M3_4

M50_4

CALCulate2

 

 

 

 

CALCulate:FEED "M3_2"

 

CALCulate2:FEED "M2_4"

ST7162

 

Figure 3.18 Post processing feed definition

3.9.1.2Specify the settings of the post processing function.

When desired, specify the settings of the post processing function to be used. The

following settings can be programmed:

 

- the filter type of the FFT function

RECTanguler HAMMing HANNing

-

the width of the low-pass filter window

3, 5, 7, .., 39, 41 points

-

the width of the differential window

3, 5, 7, .., 127, 129 points

Example:

Send CALCulate2:TRANsform:FREQuency:WINDow HAMMing

’Defines the Hamming filter for the FFT process.

3.9.1.3Enable the post processing function.

Enable the desired post processing function by using the :STATe command of the calculate function concerned. The following post processing functions are available:

STANDARD AVAILABLE:

 

 

- mathematical calculations

:MATH

 

- frequency filtering

:FILTer:FREQuency

- frequency domain transformations (FFT)

:TRANsform:FREQuency

OPTIONAL:

 

 

- histogram transformation

:TRANsform:HISTogram

- integrating traces

:INTegral

 

- differentiating traces

:DERivative (alias :DIFFerential)

Example:

 

 

Send CALCulate2:TRANsform:FREQuency:STATe ON

’Enables FFT

The post processing is automatically executed when a trace that is fed into the CALCulate function is changed. If a mathematical function is switched on, the other functions are automatically switched off.