4 - 120

COMMAND REFERENCE

TRIGger[:SEQuence[1]]:LEVel

TRIGger[:SEQuence[1]]:LEVel:AUTO

TRIGger[:STARt]:LEVel

TRIGger[:STARt]:LEVel:AUTO

Syntax:

TRIGger[:SEQuence[1]]:LEVel <NRf> MINimum MAXimum

 

TRIGger[:SEQuence[1]]:LEVel:AUTO <Boolean>

Alias:

TRIGger[:STARt]:LEVel <NRf> MINimum MAXimum

 

TRIGger[:STARt]:LEVel:AUTO <Boolean>

 

<NRf>

The trigger level expressed in volts.

 

MINimum

Selects the minimum possible trigger level.

 

MAXimum

Selects the maximum possible trigger level.

Query form: TRIGger[:SEQuence[1]]:LEVel? [MINimum MAXimum]

Alias:

TRIGger[:STARt]:LEVel? MINimum MAXimum

Response:

<NR3>

 

 

<NR3>

The trigger level in volts.

Query form: TRIGger[:SEQuence[1]]:LEVel:AUTO?

Alias: TRIGger[:STARt]:LEVel:AUTO?

Response: 0 1

0Level peak-peak off.

1Level peak-peak on.

Description:

The TRIGger:LEVel command controls the trigger level. The trigger level for the trigger source is effective only if the trigger source is INTernal 1, 2, 3 or 4. The instrument function "level-pp" is automatically switched off. If the trigger source is LINE, execution error -221, "Settings conflict" is generated at receipt of the command. Execution error -221 is also generated if the instrument cannot report the unit in volts upon receipt of the query.

The TRIGger:LEVel:AUTO switches the level peak-peak function on or off. If level peak-peak is switched off, the trigger level is automatically reactivated. If level peak-peak is switched on, the trigger level is automatically deactivated and the level range is clamped within the peaks of the signal.

Page 220
Image 220
Fluke PM-3380B TRIGgerSEQuence1LEVel NRf MINimum MAXimum, TRIGgerSEQuence1LEVelAUTO Boolean, TRIGgerSTARtLEVelAUTO Boolean