USING THE COMBISCOPE INSTRUMENTS

3 - 39

3.7 Signal Conditioning

The INPut subsystem allows you to condition the input signals, such as AC/DC/GROund coupling, input filtering, and input impedance selection.

In the digital mode, the SENSe:VOLTage<n>:RANGe:AUTO command allows you to enable autoranging of the attenuation for each of the input channels <n> separately.

@1

@2

@3

@4

INPut

SENSe

 

 

 

:FUNCtion

INPut[1]

:VOLTage[1]

:OFF

"XTIMe:VOLTage1"

 

 

 

 

[:ON]

INPut2

:VOLTage2

:OFF

"XTIMe:VOLTage2"

 

 

[:ON]

 

 

:SWEep

INPut3

:VOLTage3

:OFF

"XTIMe:VOLTage3"

 

 

[:ON]

INPut4

:VOLTage4

:OFF

"XTIMe:VOLTage4"

 

 

[:ON]

 

 

Main Time Base

 

 

ST7159

Figure 3.15 Signal conditioning

3.7.1AC/DC/ground coupling

The INPut<n>:COUPling command allows you to set the vertical input coupling at AC, DC, or GROund for each input channel separately. After a *RST command, all input channels are DC coupled.

PROGRAM EXAMPLE:

CALL

Send(0,

8,

"INPut:COUPling AC", 1)

Sets channel 1 AC coupled

CALL

Send(0,

8,

"INPut2:COUPling GROund", 1)

Sets channel 2 ground coupled

Page 58
Image 58
Fluke PM-3394B, PM-3380B, PM-3390B, PM-3370B, PM-3384B user manual Signal Conditioning, 1 AC/DC/ground coupling