USING THE COMBISCOPE INSTRUMENTS

3 - 59

3.10.3Writing data to trace memory

The TRACe command allows you to write data from the controller into a memory register. The following possibilities are available:

Write a previously read trace using the TRACe? query.

Example:

 

Send TRACe? CH3

’Queries for CH3 trace

Read <trace block>

’Reads trace data block

Send TRACe M2_3,<trace block> ’Writes data block to M2_3

The result is that trace area M2_3 is filled with the acquisition trace of channel 3.

Programming note:

The fixed command part (TRACe M2_3,) and the variable <trace block> must be sent separately. So, no EOI (End Or Identify) detection in between. Also the <trace block> must be sent without EOI detection and detection of the EOL (End Of Line) code, because the <trace block> could contain the EOL character, e.g., code 10 for CR.

Write a trace of identical constants (range = -32767 ... 32767).

Example:

Send TRACe M2_4,1028’1028 = 1024 + 4 = 0404 hex.

This command fills all memory register M2_4 locations with the constant 0404 hexadecimal for 16-bit samples, and with 04 hexadecimal for 8-bit samples.

Note: A trace can only be written to memory register space (Mi_n) and not to acquisition space (CHn).

PROGRAM EXAMPLE:

DIM response AS

STRING * 2000

 

Dimensions trace buffer

CALL Send(0, 8,

"TRACe? CH1",

1)

Requests for channel 1 trace

CALL Receive(0,

8, response$,

256)

Reads the channel 1 trace

length = IBCNT%

 

 

IBCNT = number of data bytes

CALL Send(0, 8,

"TRACe M2_3,", 0)

Sends fixed command part without EOI

CALL Send(0, 8,

LEFT$(response$,length), 0)

 

 

 

Sends variable <trace block> without EOI

CALL Send(0, 8,

"", 1)

 

Sends dummy string with EOI detection

Page 78
Image 78
Fluke PM-3394B, PM-3380B Writing data to trace memory, Send → TRACe? CH3, ’Queries for CH3 trace, Read ← trace block