Operation of the GPIB Appendix E
GPIB-1014 User Manual E-6 © National Instruments Corporation
Figure E-3. Star Configuration
Configuration RequirementsTo achieve the high data transfer rate that the GPIB was designed for, the physical distance
between devices and the number of devices on the bus are limited.
The following restrictions are typical.
• A maximum separation of 4 m between any two devices and an average separation of 2 m over
the entire bus.
• A maximum total cable length of 20 m.
• No more than 15 devices connected to each bus, with at least two-thirds powered on.