9.7 Troubleshooting Procedures

Introduction

The troubleshooting procedures are listed in numerical order as they

 

appear in Table 9-7. Each procedure lists what to do if you have that

 

particular failure and how to do it or where to find the data needed to

 

accomplish the task.

Equipment needed

Procedure #1

WARNING—SHOCK HAZARD

!

 

TROUBLESHOOTING MAY REQUIRE ACCESS TO HAZARDOUS LIVE CIRCUITS, AND SHOULD ONLY BE PERFORMED BY QUALIFIED SERVICE PERSONNEL. MORE THAN ONE SWITCH MAY BE REQUIRED TO DE-ENERGIZE UNIT BEFORE SERVICING.

You will need the following equipment in order to troubleshoot the symptoms listed in the tables that follow:

DC Milliammeter – mAdc

Calibration sources – T/C, mV, Volt, etc.

Voltmeter

Table 9-8 explains how to troubleshoot power failure symptoms.

Table 9-8 Troubleshooting Power Failure Symptoms

Step

1

2

3

4

5

What to do

Check the AC line voltage.

Make sure the chassis plugs into the rear of the case properly.

Check the system for Brown-outs, heavy load switching, etc., and conformance to installation instructions.

Change PS/Input board.

Change MCU/Output board.

How to do it

Use a voltmeter to measure the AC voltage across terminals L1 and L2 on the rear terminal panel of the controller.

Check the earth ground connection.

Withdraw the chassis and visually inspect the controller board and the inside of the case.

Refer to Section 2 - Installation.

Refer to Table 9-20.

Refer to Table 9-23.

4/00

UDC 3300 Controller Product Manual

253

Page 265
Image 265
Honeywell UDC 3300 manual Troubleshooting Procedures, Accomplish the task, Equipment needed Procedure #1, How to do it