Remote Commands

Command Examples

The following are some typical GB1400 set and query command examples:

Command

Action or Response

 

 

clock_freq?

Returns the current output clock frequency setup of

 

the Generator or the measured input clock frequency

 

of the Analyzer

 

 

word_memory?

Returns the saved word patterns in all 10 word

 

memory locations.

 

 

word_memory? 7

Returns the saved word pattern in word memory

 

location 7.

 

 

word memory 9, 16, #Haa, #H23

Sets the 16 bits of the word pattern stored at location

 

9 to the bit sequence represented by AA23

 

hexadecimal.

 

 

word_bits 8, #Q307

Sets the first eight bits of the current active pattern to

 

the bit sequence represented by 307 octal.

 

 

recall_word 7

Recalls the word pattern saved in location 7 and

 

makes it the current active pattern.

 

 

word_mem_len 6 , 16

Sets the length of the word pattern stored in location

 

6 to 16 bits.

 

 

view_angle?

Returns the current display view angle setup.

 

 

view_angle 3

Sets the view angle to 3.

 

 

rs_prompt "700"

Sets the terminal display prompt to "700".

 

 

all_mem?

Returns the word patterns saved in all 10 word

 

memory locations.

 

 

C-6