Remote Commands

rdata_delay [v]

Reference Data Delay. Sets the Analyzer REF DATA input delay to v, where v may be set in the range 0.0 nSec to 3.99 nSec in 0.1 or 0.2 nSec steps. v is expressed as a real number with an exponent of -9.

Min. Abbr.

rdata_dela

Argument

v:

0.00E-9 to 3.99E-9, in steps of 0.10E-9 or 0.2E-9

Example

rdata_delay 1.10E-9

rdata_delay?

Reference Data Delay Query. Returns the current Analyzer REF DATA input delay.

Min. Abbr.

rdata_dela?

 

Returns

<NR3>

 

Example

rdata_delay?

(command)

 

RDATA_DELAY 1.10E-9

(response)

rdata_term [neg_2v gnd ac]

Reference Data Termination. Sets the Analyzer REF DATA input termination.

Min. Abbr.

rdata_te

 

Arguments

neg_2v:

50 ohms to -2 volts.

 

gnd

50 ohms to ground.

 

ac

50 ohms via 0.01 μF capacitor to ground.

 

Example

rdata_term neg_2v

rdata_term?

Reference Data Termination Query. Returns a character string indicating the current Analyzer Reference Data Input termination setup.

Min. Abbr.

rdata_te?

 

Returns

[neg_2v gnd ac]

 

Example

rdata_term?

(command)

 

RDATA_TERM NEG_2V

(response)

GB1400 User Manual