Tektronix 071-0590-00 Display Mode Totalize, Window, or Test, Clearing Results and Starting Tests

Models: 071-0590-00

1 376
Download 376 pages 61.76 Kb
Page 110
Image 110

Reference

Display Mode: Totalize, Window, or Test

When the Analyzer is not in the Menu mode, you can select a display mode using the SELECT DISPLAY key. The display mode determines which BER and Bit Error results are shown in the main display. BER results are shown in the top, middle field of the display, while Bit Error results are shown in the top right field. Display mode also determines which results will be cleared when you press the ERROR DETECTION CLEAR key.

Procedure to Select a Results Display Mode

Press the DISPLAY SELECT key to select a different display mode. Selections are: Totalize , Window, or Test

The selected display mode will be indicated by the character in front of the BER result, as explained in the table below. Note that there are three possible Test display mode characters (T, U, and R) which further indicate which Test timing mode (Timed, Untimed, or Repeat) has been set up. Note that MODE is a parameter in the TEST Menu.

Table 3-9. How to Tell Which Display Mode is Active

Character In Front of BER Result

Indicated Result Display Mode

no character (blank)

Window

the infinity symbol ()

Totalize

 

T, U, or R

Test:

 

T = Timed mode.

 

U = Untimed mode

 

R = Repeat mode

Clearing Results and Starting Tests

To do this:

1.Press DISPLAY SELECT until the Totalize Mode results are displayed (infinity character in front of the BER result).

2.Press the CLEAR key in the ERROR DETECTION section to clear Totalize BER and Bit Error results.

Totalize Process Set Up

The Totalize measurement process is active all the time and requires no explicit setup by the operator.

GB1400 User Manual

3-33

Page 110
Image 110
Tektronix 071-0590-00 Display Mode Totalize, Window, or Test, Clearing Results and Starting Tests, Totalize Process Set Up