Remote Commands

data_term [neg_2v gnd ac]

Data Termination. Sets the Analyzer data input termination voltage to "v".

Min. Abbr.

data_te

 

Arguments

neg_2v:

50 ohms to -2 volts.

 

gnd

50 ohms to ground.

 

ac

50 ohms via 0.01 μF capacitor to ground.

 

Example

data_term ac

data_term?

Data Termination Query. Returns a character string indicating the current Analyzer data input termination setup.

Min. Abbr.

data_te?

 

Returns

[neg_2v gnd ac ]

 