Page 49
Image 49

Functional Overview

Analyzer INPUT Section

The controls shown below are used to set up the Analyzer clock and data inputs.

INPUT

DELAY

 

 

 

 

V-TERM V-THRESH

 

 

 

D-INV

 

EXT

 

 

 

 

 

 

DATA

DATA

REFERENCE DATA

CLOCK

CLOCK

 

 

(1)

 

(1)(2)

 

50 Ohm, 2V MAX

 

50 Ohm, 1.5V MAX

DELAY: Press this key to add delay to the DATA or REF DATA inputs to adjust the clock/data phase relationship. Note that the Auto_Search function will automatically set data delay to a value which provides the maximum noise immunity, that is so that the active (falling) edge of the clock falls in the middle of data bit time slots.

V-THRESH:Press this key to set the input decision threshold for the DATA and REF DATA inputs. Note that threshold does not apply when differential operation is selected. Function keys F2 and F3 are OFF when programming DATA. Function key F3 is ON when programming REF DATA.

V-TERM:Press this key to select the input terminations for the DATA,

CLOCK, or REF DATA inputs. Available selections are: (GND, -2 V, or AC). See table below.

D-INV:Press this key to select either the data non-inverted (LED off) or data inverted (LED on) mode.

EXT: This is an input for an external data reference signal.

NOTE: Use the F2 and F3 function keys to determine which input will be affected by the DELAY, V-THRESH, and V-TERM controls as follows:

F2

F3

Affected Input

Allowable Control

off

off

DATA

V-TERM, V-THRESH, DELAY

on

off

CLOCK

V-TERM only

off

on

REF DATA

V-TERM, V-THRESH, DELAY

on

on

not allowed

 

GB1400 User Manual

2-23