Tektronix 071-0590-00 user manual Functional Test

Models: 071-0590-00

1 376
Download 376 pages 61.76 Kb
Page 335
Image 335

Performance Verification

Functional Test

To functionally test the GB1400, connect the Generator to the Analyzer and confirm their correct operation as described below.

Note

In these procedures, the Generator and Analyzer are returned to their default settings. The word memories will be reset. If you have entered word patterns that you do not want to lose, use a GPIB or RS-232C controller to save them before beginning these procedures.

qStep 1: Make connections appropriate to the options installed as follows:

Connections for Functional Test

Option

From

To

Comment

Standard Instrument

TX Clock

RX Clock

50 Ohm, SMA coax

 

TX Not Clock

RX Not Clock

50 Ohm, SMA coax

 

TX Data

RX Data

50 Ohm, SMA coax

 

TX Not Data

RX Not Data

50 Ohm, SMA coax

75 Ohm Option

TX Clock

RX Clock

50 Ohm, SMA coax.

 

TX Not Clock

RX Not Clock

Use SMA to BNC

 

TX Data

RX Data

adapters and 75 Ohm

 

BNC coaxial cables

 

TX Not Data

RX Not Data

 

 

See Figure 1 for Setup of Functional Test.

qStep 2: Reset both the Generator and Analyzer to their factory default settings. Press the ‘CLEAR’ key in the PATTERN section, the ‘MSB(1)’ key and the ‘VIEW ANGLE’ key, and hold down all three keys while cycling the power.

The message DEFAULT SETTINGS should appear briefly on the display of each unit, followed by the GB1400 logo, before the functional display appears.

qStep 3: At the Generator, verify that error injection is off (you should see ERR OFF on the display). If error injection is on, press the ‘SINGLE’ key in the ERROR INJECT section to turn it off.

GB1400 Acceptance Test

E-3

Page 335
Image 335
Tektronix 071-0590-00 user manual Functional Test