Specifications

Auxiliary Signals: Pattern Sync, Clock, Data, Error and Threshold

Level

250 mVp-p into 50 Ohms, 500 mV into Hi-Z

Impedance

50 Ohms

Data Monitor

Latched Input Data

Clock Monitor

Buffered Input Clock

Pattern SYNC

1-bit wide pulse per frame

Error Inhibit

Rear panel, ECL

Error Output

Rear panel, ECL

AUX

Rear panel, Data Threshold output

 

Note: Connect this output to DATA BAR (Not DATA) for

 

single-ended operation.

Connectors

SMA

Synchronization

 

Auto Search

 

Unit automatically finds the Data Threshold , the Clock/Data input timing phase delay, the input Data Pattern (PRBS or WORD mode), and Data Polarity.

There are two modes to find the Data Delay

FAST - A quick method using the Clock/Data phase indicator.

BER - A slower method which uses the signal's bit error rate.

Either method will make available the width of the Data Eye (if possible). The BER method allows user control over:

Data Sample Size (10E-4 to 10E-11). This is the number of data bits sampled at each Delay setting used to determine the center of the Data Eye.

Bit Error Rate Threshold (10E-3 to 10E-10). This is the threshold used to determine which Delay settings are part of the Data Eye Crossing.

Manual Mode

User selects parameters, then unit attempts to synchronize on the input data pattern.

Disable Mode

SYNC circuitry disabled

GB1400 User Manual

A-7

Page 171
Image 171