Tutorial

Set DATA

Verify these results on the Analyzer

Comments

THRESHOLD

 

 

to this value

 

 

 

Press the ERROR DETECTION and ERROR

Data signal not detected. Selected

-0.50 volts

HISTORY CLEAR KEYS”, then confirm:

threshold is below the minimum level of

SYNC LOSS and BIT LED’s are ON

your data signal.

 

 

 

SYNC LOCK LED is OFF

 

 

ERROR RATE display shows “NO

 

 

DATA” or a 50% Error Rate.

 

 

 

 

 

Press the ERROR DETECTION and ERROR

You are starting to detect the data signal.

Approximately

HISTORY CLEAR KEYS”, then confirm:

Selected threshold is near the minimum

 

level of your data signal. Data threshold

-0.30 volts

SYNC LOSS should turn OFF

is NOT CORRECT and you should expect

 

 

to

BIT and PHASE LED’s should turn ON

BER errors.

 

 

-0.25 volts

 

 

 

SYNC LOCK LED should turn ON

 

 

 

 

 

Press the ERROR DETECTION and ERROR

Data signal fully detected. Selected

+0.00 volts

HISTORY CLEAR KEYS”, then confirm:

threshold is at an optimum value for the

SYNC LOSS should turn OFF

input data signal. This is typically =

 

1/2*[max level - min level].

 

BIT and PHASE LED’s should turn OFF

Signal levels detected above this

 

SYNC LOCK LED should turn ON

threshold are considered a logical “1”

 

and signal levels below this threshold are

 

ERROR RATE display shows NO BER

considered a logical “0”.

 

 

 

ERRORS equivalent to a rate of

 

 

0.0E-9 or better.

 

 

 

 

 

Press the ERROR DETECTION and ERROR

Data signal not detected. Selected

+1.00 volts

HISTORY CLEAR KEYS”, then confirm:

threshold is above the maximum level of

SYNC LOSS and BIT LED’s are ON

your data signal.

 

 

 

SYNC LOCK LED is OFF

 

 

ERROR RATE display shows “NO

 

 

DATA” or a 50% Error Rate

 

2-36

GB1400 User Manual

Page 62
Image 62
Tektronix 071-0590-00 Data signal not detected. Selected, Threshold is below the minimum level, Your data signal