Remote Commands

Examplesclock_off_up clock_off_dn

clock_off_up [v] and clock_off_dn [v]

Clock Offset Up, Clock Offset Down. This command pair is used to increment or decrement the current output clock signal baseline offset by a specified amount [v] volts.

Min. Abbr.

clock_off_u

 

clock_off_d

Arguments

v: <NR2>, in the range 0.05 to 3.00 in 0.05 v steps.

 

in the range 0.05 to 3.80 in 0.05 v steps (PECL option)

Example

clock_off_up 0.10

 

clock_off_dn 0.10

clock_offset [v]

Clock Offset. Sets the baseline offset of the output clock signal to [v] volts.

Min. Abbr.

clock_offs

Arguments

v:

<NR2>, in the range -2.00 to 1.00 in steps of 0.05.

 

 

in the range -2.00 to 1.80 in 0.05 v steps (PECL option)

Example

clock_offset -0.75

clock_offset?

Clock Offset Query. Returns the current value of the output clock signal baseline offset.

Min. Abbr.

clock_offs?

 

Returns

<NR2>

 

Example

clock_offset?

(command)

 

CLOCK_OFFSET -0.75

(response)

data_amp_up and data_amp_dn

Data Amplitude Up/Down (Default). This command pair is used to increment or decrement the current amplitude of the Generator data output by a default amount of 0.05 volts.