Remote Commands

rdata_thr_up and rdata_thr_dn

Reference Data Threshold Up/Down (Default). This command pair is used to increment (rdata_thr_up) or decrement (rdata_thr_dn) the current REF DATA threshold by the default increment of 0.05 volts.

Min. Abbr.

rdata_thr_u

 

rdata_thr_d

Argument

none

Examples

rdata_thr_up

 

rdata_thr_dn

rdata_thr_up [v] and rdata_thr_dn [v]

Reference Data Threshold Up/Down. This command pair is used to increment (rdata_thr_up) or decrement (rdata_thr_dn) the current REF DATA threshold by v, where v is in <NR2> format and may be set in the range 0.05 to 2.50 Volts in 0.05 V steps.

Min. Abbr.

rdata_thr_u

 

rdata_thr_d

Argument

v:

<NR2>, 0.05 to 2.50 volts, in 0.05 V steps.

Examples

rdata_thr_up 0.10

 

rdata_thr_dn 0.50

rdata_thres [v]

Reference Data Threshold. Sets the threshold voltage for the Analyzer REF DATA input to v, where the allowed range for v depends on the current input clock termination.

Min. Abbr.

rdata_thre

Arguments

v:

-1.50 to 1.00 in 0.05 v steps, for terminations

 

 

GND, AC

 

 

-2.50 to 0.00 in 0.05 v steps, for termination -2V

Example

rdata_thres 1.00

C-64

GB1400 User Manual

Page 282
Image 282