WINDOW interval "W" is set in terms of bits or time in using WINDOW menu.

Slip time "S" is effectively 200 ms in display or 1 sec. in End-of-Window reports

CLEAR

W

W

W

S

W

W

Figure 3-5. The WINDOW Measurement Process

How TEST Results Are Measured

Unlike Totalize and Window results, the accumulation of Test results can be started or stopped by pressing the CLEAR key. That is, the CLEAR key in effect becomes a test start/stop key. While a test is "stopped", all current Test results are frozen. When the CLEAR key is pressed to start a new test, all current Test results are saved as previous Test results before current result registers are cleared.