Remote Commands

data_thres?

Data Threshold Query. Returns the current threshold voltage for the Analyzer DATA input.

Min. Abbr.

data_thre?

 

Returns

<NR2>

 

Example

data_thres?

(command)

 

DATA_THRES -1.50

(response)

rdata_del_up and rdata_del_dn

Reference Data Delay Up/Down (Default). This command pair is used to increment (rdata_del_up) or decrement (rdata_del_dn) the current REF DATA delay by the default increment of 0.1 nSec.

Min. Abbr.

rdata_del_u

 

rdata_del_d

Arguments

none

Examples

rdata_del_up

 

rdata_del_dn

rdata_del_up [v] and rdata_del_dn [v]

Reference Data Delay Up/Down. This command pair is used to increment

(rdata_del_up) or decrement (rdata_del_dn) the current REF DATA delay by v, where v is in <NR3> format and may be set in the range 0.1 nSec to 3.99 nSec in 0.1 nSec steps.

Min. Abbr.

rdata_del_u

 

rdata_del_d

Argument

v:

<NR3>, 0.10E-9 to 3.90E-9 in 0.10E-9 steps.

Examples

rdata_del_up 1.00E-9

 

rdata_del_dn 1.00E-9

C-62

GB1400 User Manual

Page 280
Image 280
Tektronix 071-0590-00 user manual Datathres?, Rdatadelup and rdatadeldn, Rdatadelup v and rdatadeldn