Tektronix 071-0590-00 user manual Winmode bitssec, Winmode?, Winprev currentprevious, Winprev?

Models: 071-0590-00

1 376
Download 376 pages 61.76 Kb
Page 299
Image 299

Remote Commands

win_mode [bitssec]

Windows Mode. Determines whether BER window length is determined by the windows bits (pattern length) parameter or the windows seconds (time length) parameter.

Min. Abbr.

win_m

 

Arguments

bits:

selects the windows bits length parameter

 

sec:

selects the windows seconds length parameter.

Example

win_mode bits

win_mode?

Windows Mode Query. Returns the current BER window length mode.

Min. Abbr.

win_m

 

Returns

[bitssec]

 

Example

win_mode?

(command)

 

WIN_MODE BITS

(response)

win_prev [currentprevious]

Window Previous. Sets the WIN_PREV parameter to current or previous. This parameter determines whether Analyzer responses to win_bits?, win_error?, win_rate?, and win_time? queries will be based on the current window interval or the previous window interval.

Min. Abbr.

win_p

 

Arguments

current

selects the current window interval

 

previous

selects the previous window interval

Example

win_prev current

win_prev?

Window Previous Query. Returns the current value of the WIN_PREV parameter.

Min. Abbr.

win_p

 

Returns

[currentprevious]

 

Example

win_prev?

(command)

 

WIN_PREV CURRENT

(response)