data_thr_u data_thr_d

Remote Commands

data_thr_up and data_thr_dn

Data Threshold Up/Down (Default). This pair of commands increments (data_thr_up) or decrements (data_thr_dn) the current Analyzer DATA input threshold by the default increment of 0.05 volts.

Min. Abbr.

data_thr_u

 

 

data_thr_d

 

Arguments

none

 

Examples

data_thr_up

(increments threshold by 0.05 volts)

 

data_thr_dn

(decrements threshold by 0.05 volts)

data_thr_up [v] and data_thr_dn [v]

Data Threshold Up/Down. This pair of commands increments (data_thr_up) or decrements (data_thr_dn) the current Analyzer DATA input threshold by an amount v.

Min. Abbr.

Arguments

v:

0.05 to 2.50 in 0.05 V steps

Examples

data_thr_up 0.10

(increments threshold by 0.1 volt)

 

data_thr_dn 2.00

(decrements threshold by 2 volts)

data_thres [v]

Data Threshold. Sets the threshold voltage for the Analyzer DATA input to v, where the allowed range for v depends on the current DATA input termination.

Min. Abbr.

data_thre

Arguments

v:

-1.50 to 1.00 in 0.05 v steps, for input terminations

 

 

GND, AC

 

 

-2.50 to 0.00 in 0.05 v steps, for input terminations -2V

Example

data_thres -1.50

GB1400 User Manual