Reference

As shown in the next figure, Test results may be calculated using one of three timing modes: Timed, Untimed, or Repeat.

T EST Measurement Process

C LEAR

U ntimed T EST Process:

U ntimed

Intervals started andMeasurement Interval stopped manually by operator.

TEST interval "T" set in terms of time (hrs:min:sec) in TEST menu.

C LEAR

Timed TEST Pr ocess: Interv als started by operator, stopped auto - matically at end of interval T.

C LEAR

R epeat Timed TEST Process: Process started by operator, intervals of length T stopped and re -s tarted automatically.

 

C LEAR

C LEAR

T

T

T

T T T T T

Figure 3-6. The TEST Measurement Process

Timed Tests. When the Timed mode is selected, the Analyzer will automatically stop accumulating Test results after the test interval (specified by the user) has elapsed. All Test results will be frozen until the CLEAR key is pressed to begin a new test.

Repeat (Timed) Tests. When the Repeat mode is selected, the Analyzer will automatically stop and then restart a test after a user specified test length has elapsed. This process will continue until the user presses the CLEAR key to manually stop Test result accumulation.

Untimed Tests. When the Untimed mode is selected, the Analyzer will continue to collect Test results until the CLEAR key is pressed. At this point Test results will remain frozen until the CLEAR key is pressed again to start a new test.

3-32