Chapter 4, Description of Fields

Fields

Operating Considerations

A sequence of access probes is transmitted as part of the mobile station’s attempt to gain access to a CDMA system. To determine a mobile station’s access probe power, select Acc Prb Pwr from the list of power measurement choices. Press the Config key to access the CDMA CELL SITE CONFIGURATION screen. Select Page in the Call limit field. This will prevent a call from going to the Connected state, resulting in the mobile station transmitting all access probes in its pre- defined sequence. Several field settings affect the number of access probes and power levels of each access probe. These fields are:

"RF Power" field on page 371 "Sctr A Pwr" field on page 380 "Sctr B Pwr" field on page 381 "Nom Power" field on page 312 "Nom Pwr Ext" field on page 312 "Init Power" field on page 274 "Power Step" field on page 342 "Num Step" field on page 317

"Max Req Seq, Max Rsp Seq" field on page 294

Screens Where Field is Present

CDMA CALL CONTROL

CDMA CELLULAR MOBILE TRANSMITTER TEST

CDMA TRANSMITTER POWER RANGE TEST

150