Agilent Technologies E8285A manual AFGen1 To, AFGen2 Freq, This field has two fields

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 158
Image 158

Chapter 4, Description of Fields

Fields

AFGen1 To

This field has two fields:

the upper field sets the destination port for Audio Frequency Generator 1

FM -RF Generator FM modulator

AM -RF Generator AM modulator

Audio Out -AUDIO OUT connector on front panel of Test Set

the lower field sets the:

FM modulation deviation if the upper field set to FM

AM modulation depth if the upper field set to AM

amplitude of audio signal (volts RMS) at the AUDIO OUT connector if upper field is set to Audio Out

For testing mobile stations, the upper field is normally set to FM, and the lower field set to the desired FM deviation in kHz. See “AF Generator 1” in the GPIB Command Syntax chapter of the E8285A Condensed Programming Reference Guide for programming command syntax.

Operating Considerations

This is a priority control field. Accessing the RX TEST or TX TEST screen overrides any changes made to this field in other screens.

Screens Where Field is Present

ADJACENT CHANNEL POWER

ANALOG MEAS

DUPLEX TEST

RX TEST

RF GENERATOR

AFGen2 Freq

This Audio Frequency Generator control sets the frequency for the second audio frequency sinewave generator.

Screens Where Field is Present

RF GENERATOR

RX TEST

158