Chapter 1, Description of Keys

Keys

Ref set

Press and release the Shift key, then the Increment÷10 key to access the reference set function. This function is used to enter or remove a measurement reference for relative AF and RF measurements.

Programming Ref set

The GPIB command :REFerence is used to select this data function programmatically.

To Turn Measurement Reference Points ON and OFF. Use the :REFerence:STATe <boolean> commands to turn measurement reference points ON and OFF. The <boolean> parameter can be a 1 or ON to turn measurement reference points on, and a 0 or OFF to turn measurement reference points off.

Syntax

:REFerence:STATe <ON> or <1> :REFerence:STATe <OFF> or <0>

Example

OUTPUT 714;"MEAS:RFR:POW:REF:STAT ON"

This turns the measurement reference point for the TX power measurement

ON.

To Query the State of Measurement Reference Points. Use the :REFerence:STATe? commands to query the state of a measurement reference point. The query returns a 1 if a measurement reference points is ON, and a 0 if a measurement reference points is OFF.

Syntax

:REFerence:STATe?

Example

OUTPUT 714;"MEAS:RFR:POW:REF:STAT?"

ENTER 714;Meter_on_off ! returns a 1 (ON) or 0 (OFF)

This queries the state of the measurement reference point for the TX power measurement.