Chapter 1, Description of Keys

Keys

NOTE:When querying a field setting or measurement result through GPIB, the Test Set always returns numeric values in GPIB units or attribute units, regardless of the field’s current display units setting. Refer to “To Specify Units-of-Measure for GPIB Data Transfer” in the Operating Overview chapter of the E8285A User’s Guide .

To Set the Increment Mode. Use the :INCRement:MODE commands to set the increment mode to linear or logarithmic.

Syntax

:INCRement:MODE <LOGarithm or LINear>

Example

OUTPUT 714;"RFG:FREQ:INCR:MODE LOG"

This sets the increment mode for the RF Generator’s frequency to logarithmic.

To Query the Increment Mode. Use the :INCRement:MODE? commands to query the increment mode.

Syntax

:INCRement:MODE?

Example

OUTPUT 714;"RFG:FREQ:INCR:MODE?"

ENTER 714;Mode$ ! returns LIN or LOG

This queries the increment mode of the RF Generator’s frequency.

To Set the Increment Value Display Units. Use the :INCRement:DUNits commands to set the units-of-measure used to display the increment value. Refer to “To Specify Units-of-Measure for CRT Display” in the Operating Overview chapter of the E8285A User’s Guide for description of display units.

Syntax

:INCRement:DUNits <disp_units>

Example

OUTPUT 714;"RFG:FREQ:INCR:DUN KHZ"

This sets the increment value’s display units to kHz for the RF Generator’s frequency.