Agilent Technologies E8285A manual RF Level Offset

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 369
Image 369

Chapter 4, Description of Fields

Fields

RF In

This field is used to indicate reverse channel path losses or gains between the mobile station and the RF IN/OUT port.

Enter a positive value to indicate a gain (such as amplifier gain).

Enter a negative value to indicate a loss (such as cable loss).

The RF In field is displayed only when RF In is selected in the Input Port field.

GPIB Example

“CONF:OFL:RF:IN -1”

enters a path loss of 1 dB between the RF IN/OUT port (input path only) and the mobile station.

The value entered into this field is applied only when the RF Level Offset field is set to On.

See Also

RF Level Offset field description, on page 369

Screens Where Field is Present

CONFIGURE

RF Level Offset

This field enables/disables the RF level offsets entered in the RF In/Out,

Duplex Out, and Antenna In fields below it.

When set to On, the RF Generator’s amplitude and RF Analyzer’s power measurement are offset by the values entered in these fields.

When set to Off, the values in these fields are ignored.

See Also

Antenna In field description, on page 167

Duplex Out field description, on page 242

RF In field description, on page 369

Screens Where Field is Present

CONFIGURE

369