Chapter 4, Description of Fields

Fields

Sat Tol

This field is displayed only for NAMPS system type.

This field is used to select the Sat tolerance. When set to Narrow the Sat limit is one of the three valid Sat tones ± 10Hz. When set to Wide the Sat limit is defined at 0 to 7000 Hz. When operating in noisy environments the narrow Sat limit may not allow for correct measurements and the ability to alter the limit to encompass a wider range provides for testing flexibility.

The :STOLerance command is used to control this field.

The query form (that is, :STOLerance?) can be used to query the current field value.

Syntax

:STOLerance <‘><NARROW/WIDE><’>

:STOLerance?

Example

OUTPUT 714;"CALLP:AMPS:STOL ‘NARROW’"

OUTPUT 714;"CALLP:AMPS:STOL?"

ENTER 714;Stol_set$

Screens Where Field is Present

CALL CONTROL

CALL CONFIGURE