Agilent Technologies E8285A manual Chan Power Meas Intrvl, Sets the measurement interval to 1 ms

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 198
Image 198

Chapter 4, Description of Fields

Fields

Operating Considerations for the Channel Power Measurement

The cross-calibration process is performed by selecting the Power Meas, Calibrate field. Refer to “Calibrating the Test Set” in the Calibrating the Test Set chapter of the E8285A Application Guide. Channel power measurements must be calibrated whenever any of the following conditions occur:

Uncal annunciator is flashing

5 °C change in operating temperature

Test Set power cycle

Daily

The Average Power measurement, which is used during cross-calibration, is automatically zeroed during the calibration procedure.

No special measurement triggers are required for Channel Power measurements.

See Also

Avg Power field description, on page 178

Screens Where Field is Present

CDMA CALL CONTROL

CDMA TRANSMITTER POWER RANGE TEST

CDMA CELLULAR MOBILE TRANSMITER TEST

Chan Power Meas Intrvl

This field allows you to set the interval for the Channel Power measurement. This setting is only valid with Service Option 2 or Service Option 9, and during Full Rate.

GPIB Examples

“CDManalyzer:CHANpower:INTerval 1”

sets the measurement interval to 1 ms.

Screens Where Field is Present

CDMA TX

198

S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp

Page 198
Image 198
Agilent Technologies E8285A manual Chan Power Meas Intrvl, Sets the measurement interval to 1 ms