Chapter 4, Description of Fields

Fields

Release

This field is used to terminate an active voice channel connection with the mobile station. When the Release field is selected, a mobile station control message with a release order is sent to the mobile station on the forward voice channel. A mobile station must be actively connected on a voice channel to the Test Set (that is, the Connect annunciator must be lit) before you can attempt to send a release order to the mobile station.

The :RELease command is used to control this field.

There is no query form of the :RELease command.

Syntax

:RELease

Example

OUTPUT 714;"CALLP:REL"

Screens Where Field is Present

CALL CONTROL

Res BW

The resolution bandwidth field selects the IF filter used when measuring the power of the carrier and the adjacent channels.

Operating Considerations

Using a narrower bandwidth filter (300 Hz) slows the measurement, but rejects carrier leakage and out of channel spurs. Using a wider bandwidth filter

(1 kHz) speeds measurements, but may allow unwanted spurs and carrier leakage to be integrated into the measurement when measuring at the edges of the selected channel bandwidth.

Screens Where Field is Present

ADJACENT CHANNEL POWER

Reset

See the Cont/Single field description, on page 211

Screens Where Field is Present

OSCILLOSCOPE

362

S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp

Page 362
Image 362
Agilent Technologies E8285A manual Res BW, Reset