Chapter 4, Description of Fields

Fields

Srch Win R

This field allows entry of search-window size for the remaining set.

GPIB Example

“DISP CCON;CDMA:CELL:CONFigure:RWIN 5“

sets search-window R to 5.

Operating Considerations

The mobile station searches for the strongest pilot channel signal and usable multipath components of the pilot channel signals included in the remaining set. The search-window limits the range of PN sequence offsets that the mobile station searches.

The range of values for the searcher window size is 0 to 15.

0 = 4 PN chips (minimum)

5 = 20 PN chips (default)

15 = 452 PN chips (maximum)

Screens Where Field is Present

CDMA CELL SITE CONFIGURATION

SSD_A=0 (annunciator)

This annunciator, when lit, indicates that SSD (Shared Secret Data) in the Test Set is 0. SSD_A=0 is the default value for SSD_A. Entering a new A-Key initializes SSD_A to 0. Performing a successful SSD Update causes this annunciator to be turned off.

See Also

Call Status field description, on page 190

SSD Update field description, on page 409

Screens Where Field is Present

CDMA AUTHENTICATION

406

S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp

Page 406
Image 406
Agilent Technologies E8285A manual Srch Win R, SSDA=0 annunciator, Sets search-window R to