Chapter 4, Description of Fields

Fields

Pwr Step - MIN_PWR_CNTL_STEP. Minimum closed loop power control step size supported by mobile station. Possible values are: 1 dB, 0.5 dB, or 0.25 dB (message parameters 0, 1, 2 respectively). Displayed when TIA/EIA-95B is selected in the

"Protocol" field on page 346

Called Num - The number dialed on the mobile station handset during the last mobile station originated call.

*Clr All* - Selecting this Choice clears all MSDatabase values, allowing verification that information currently displayed is not from a previous registration.

GPIB Example

“DISP CCNT;CDMA:MOBile:DAT ‘ESN’;:CDMA:MOBile:DAT:VALue?“

sets the desired indicator to “ESN” and returns the 4-digit value for ESN.

Operating Considerations

These indicators are updated each time a mobile registration occurs.

The database is cleared whenever the protocol is changed. (See "Protocol" field on page 346 for more information about changing protocol.)

Screens Where Field is Present

CDMA CALL CONTROL

CDMA SHORT MESSAGE SERVICE

CDMA AUTHENTICATION

MS FER

This field displays the mobile station’s reported FER (frame error rate). The number displayed is a ratio of bad frames to total frames.

The mobile station can be commanded through the Test Set to change its periodic FER reporting intervals through the MS FER Report Interval field.

GPIB Example

“DISP CFER;CDMA:MOBile:FER?“

returns integer value of mobile station reported FER percentage.

See Also

MS FER Report Interval field description, on page 308 by # errors field description, on page 187

by # frames field description, on page 188

Screens Where Field is Present

CDMA MOBILE REPORTED FER