Chapter 4, Description of Fields

Fields

Pwr Lvl: -

The Pwr Lvl: field is divided into two fields:

Left-Hand Field

The left-hand field displays the mobile station’s output power level assignment for the voice channel currently being used by the Test Set and the mobile station.

A numeric value is only displayed when a mobile station is actively connected on a voice channel (that is, the Connect annunciator is lit). A “-” is displayed if a mobile station is not actively connected on a voice channel.

This is a read only field.

The :AVCPower? command is used to query the contents of the left-hand subfield.

There is no command form of the :AVCPower? command.

Syntax

:AVCPower?

Example

OUTPUT 714;"CALLP:AVCP?"

ENTER 714;Active_vc_pwr$

Right-Hand Field

The right-hand subfield (highlighted field) is used to enter the Voice Mobile Attenuation Code (VMAC). The VMAC determines the mobile station power level to be used on the designated voice channel (the channel number entered into the Chan: right-hand subfield).

The :VMACode command is used to control the right-hand subfield.

The query form of the command (that is, :VMACode?) can be used to determine the current VMAC setting.

Syntax

:VMACode?

:VMACode <integer number 0 to 7>

Example

OUTPUT 714;"CALLP:VMAC 3"

OUTPUT 714;"CALLP:VMAC?"

ENTER 714;Vmac_setting