Chapter 4, Description of Fields

Fields

Steps

This field allows entry of the number of power control bits that the Test Set will send to the mobile station through the power control subchannel when the Execute field is selected.

GPIB Example

“DISP CTXR;CDMA:PCONtrol:STEPs 100”

selects 100 power control bits to be sent on the power control subchannel.

Operating Considerations

Range of values: 1 through 150

See Also

Closed Loop Pwr Cntl field description, on page 203,

Change field description, on page 193

Execute (Closed Loop Power Control Change) field description, on page

251

Screens Where Field is Present

CDMA TRANSMITTER POWER RANGE TEST

Step Size

This field is displayed on the OPEN LOOP TIME RESPONSE screen when “Main” is selected in the "Controls (Open Loop Time Response)" field on page 212. This field allows entry of the step increase in the Test Set’s output power to be used in the next execution of the open loop power control test.

-25 dB -20 dB -15 dB -10 dB +10 dB +15 dB +20 dB +25 dB

GPIB Example

“CDMA:OLTR:STEP ‘+20 dB’”

selects a step increase of +20 dB when the next time response to open loop power control test is executed.