Chapter 4, Description of Fields

Fields

Ref Level

Reference level sets the amplitude reference level for the top line of the display.

All signals displayed are referenced to this line. This field is displayed when the

Controls field is set to Main.

GPIB Example

“CSP:RLEV 0”

sets the reference level to 0 dBm.

Operating Considerations

The unit-of-measure for the reference can be changed as needed. For instance, 0 dBm, 0.224 V, 107.0 dBmV, and 0.00100 W can all be used to represent the same level.

Screens Where Field is Present

SPECTRUM ANALYZER

CDMA REVERSE CHANNEL SPECTRUM

Ref Level (CDMA Swept Power Measurement)

This field sets the amount of the high value on the 40-dB vertical range of the display.

GPIB Example

“CDMA:SWEPtpower:RLEVel ‘15’”

sets the reference level high value.

Operating Considerations

Enter a value here that represents the reference level of the mobile station before it is ramped. This will allow the DSP to autorange and capture the trace correctly. The selectable range is +35 to 25 dBm. The default value is +5 dBm.

This value is also used to set the input path gain on the Auxiliary menu Input

Atten field.

Screens Where Field is Present

CDMA SWEPT POWER MEASUREMENT (When the Controls field is set to

Main)