Agilent Technologies E8285A manual Rgstr NID, Queries the RF Power level

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 371
Image 371

Chapter 4, Description of Fields

Fields

RF Power

This field displays Io, the total power spectral density of Sector A power, Sector B power and AWGN.

GPIB Example

“CDMA:RFP?”

queries the RF Power level.

Operating Considerations

RF Power is controlled by settings in the Sector A Power, Sector B Power, and AWGN fields. It is not directly settable.

Screens Where Field is Present

CDMA GENERATOR CONTROL

CDMA MOBILE REPORTED FER

CDMA REVERSE CHANNEL SPECTRUM

CDMA CELLULAR MOBILE TRANSMITTER TEST

Rgstr NID

This field allows entry of the cell site’s alternate network identification (NID), used only for registration.

GPIB Example

“DISP CCON;CDMA:CELL:CONFigure:RNID 0“

sets the register network identification of the simulated CDMA cell site to 0.

Operating Considerations

The Test Set will send the Rgstr NID and Rgstr SID during the registration process (when the Register field is selected) in an attempt to cause a zone-based registration to occur. Immediately following the zone-based registration, another registration is performed using the NID and SID in the Network ID and System ID fields. The latter values will determine if the mobile station is roaming.

Screens Where Field is Present

CDMA CELL SITE CONFIGURATION

371

Page 371
Image 371
Agilent Technologies E8285A manual Rgstr NID, Queries the RF Power level