Chapter 4, Description of Fields

Fields

Ext NGHB

This field allows the user to select whether to enable the custom neighbor list, the Sc Priority field, and the NGHB Fq (Ch) field.

GPIB Example

“CDMA:CELL:CONF:EXNB ‘On’”

enables or disables the custom neighbor list and associated fields.

Operating Considerations

This is an immediate action toggle field. If it is set to On, the Sc Priority field and the NGHB Fq (Ch) field appear on the screen, and the custom neighbor list is enabled. If it is set to Off, the fields do not appear and the list is disabled.

The default is Off.

The field appears only if a valid protocol is selected in the Protocol field. The valid protocols are:

TIA-EIA-98-B

ARIB-53

J-STD-008

Kor PCS

Screens Where Field is Present

CDMA CELL SITE CONFIGURATION

The time response to open loop power control tests the mobile station’s response to a step change in the Test Set’s output power.

GPIB Example

“STATus:OLTR:EVENt?”

queries the Open Loop Time Response status event register. If bit 1 is set, the last gated power test passed. If bit 2 is set, the last gated power test failed.

When a mobile station is operating in variable data rate transmission mode, the RF power is gated on during randomized 1.25 ms periods. During a CDMA Gated Power measurement the Test Set analyzes an ensemble of gated-on periods, referred to as bursts. The number of bursts used is controlled by the "Num of Bits"

field on page 316