Chapter 4, Description of Fields

Fields

IF Filter

This field selects the desired IF filter bandwidth for modulated signals being analyzed.

Screens Where Field is Present

TX TEST

DUPLEX TEST

RF ANALYZER

If Unit-Under-Test Fails

This field selects what happens if the measured results do not meet criteria for passing the current test. The underlined entry is the active choice.

Continue continues with testing even though a test in the test sequence fails to meet its test specification limits. When this occurs, an error is listed on the test-results print- out, is displayed on the CRT, or both.

Stop stops the test when a failure occurs and requires operator intervention before test- ing proceeds.

Screens Where Field is Present

TESTS (Execution Conditions)

Init Power

This field allows entry of the initial power offset for system access (INIT_PWR).

GPIB Example

“DISP CCON;CDMA:CELL:CONFigure:INITial:POWer 0“

sets INIT_PWR to 0.

Operating Considerations

This field specifies a correction factor the mobile station adds to the estimated open-loop output power for setting the power level of the first (initial) access probe in an access probe sequence.

Screens Where Field is Present

CDMA CELL SITE CONFIGURATION

274