Image 292

Chapter 4, Description of Fields

Fields

Mask Type

This field is used to select the power versus time mask used during gated power measurements to determine pass/fail results.

Original selects the mask specified by test standards IS-98A and TIA/EIA-98B.

TIA/EIA-98-Cselects the mask specified by test standard TIA/EIA-98C.

Screens Where Field is Present

CDMA GATED POWER (when Display is selected from the Controls field.)

Max Frames

This field sets the maximum frame count for each FER test.

GPIB Example

“DISP CRXT;MEASure:CDManalyzer:FERasure:MAXimum:FRAMes 1000”

sets the maximum number of frames for FER measurements to 1000.

Operating Considerations

FER measurements will be made on the number of frames entered in this field unless the test terminates because a confidence limit is reached.

Range of values: 25 through 10000000

See Also

Confidence field description, on page 207

Screens Where Field is Present

CDMA CELLULAR MOBILE RECEIVER TEST

292

S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp

Page 292
Image 292