Agilent Technologies E8285A Bw=, By # errors, This field is coupled to the Span field’s setting

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 187
Image 187

Chapter 4, Description of Fields

Fields

BW=

This field displays the resolution bandwidth of the CDMA spectrum analyzer.

Operating Considerations

This field is coupled to the Span field’s setting.

Range of Values

1 kHz

3 kHz

30 kHz

Screens Where Field is Present

CDMA REVERSE CHANNEL SPECTRUM

by # errors

When this field is On, the mobile station reports FER at least as often as the number of errors entered in the numeric entry portion below On/Off.

The numeric entry portion of the by # errors field is field is not displayed when this selection is Off.

GPIB Example

“DISP CFER;CDMA:MOBile:FERasure:REPort:BY:ERRors:STATe: ’ON’“

causes the mobile station to report FER whenever the number of errors counted by the mobile station reaches the value entered by the command below.

“DISP CFER;CDMA:MOBile:FERasure:REPort::BY:ERRors 5“

causes the mobile station to report FER if the FER count reaches 5 within the num- ber of frames specified in the MS FER Report Interval.

Operating Considerations

If by # frames is On, it is possible that mobile station reported FER will occur more frequently than the error count specified in by # errors.

Range of values: 1 through 15

See Also

by # frames field description, on page 188

Screens Where Field is Present

CDMA MOBILE REPORTED FER

187

Page 187
Image 187
Agilent Technologies E8285A manual Bw=, By # errors, This field is coupled to the Span field’s setting, Range of Values