Chapter 4, Description of Fields

Fields

Connect (annunciator)

When lit, the Connect annunciator indicates that the mobile station is connected to the Test Set on a voice channel.

NOTE:When the CALL CONTROL screen is displayed and the Call Processing Subsystem is in the connected state (Connect annunciator is lit), the Test Set constantly monitors the mobile station’s transmitted carrier power. If the power falls below 0.0005 Watts the Test Set will terminate the call and return to the active state (Active annunciator is lit). The mobile station’s transmitted carrier power is monitored on all Call Processing Subsystem screens except the ANALOG MEAS screen.

If the Test Set is in the connected state (Connect annunciator is lit) and a change is made to the Cntl Channel field the connected state will be lost.

The state of the Connect annunciator is reflected in the Call Processing Status

Register Group Condition Register bit 5.

See “Call Processing Status Register Group“ in the Status Reporting chapter of the E8285A User’s Guide for further information.

The Connect annunciator is not programmable.

Screens Where Field is Present

CALL CONTROL

CALL BIT

CALL DATA