Chapter 4, Description of Fields

Fields

Power Cntl Step Size

This field is displayed only when the "Protocol" field on page 346 is set to TIA/

EIA-95B.

This field allows entry of the following closed loop power control step sizes:

1 dB

0.5 dB

0.25 dB

Operating Considerations,

The power control step size entered in this field must not be less than the minimum step size supported by the mobile station under test.

The value entered in this field will be sent to the mobile station in the Power Control Message PWR_CNTL_STEP field during call setup or while on a traffic channel (call connected).

The mobile station’s minimum step size, MIN_PWR_CNTL_STEP, is retrieved by the Test Set during registration and displayed in the "MS Database" field on page 306 under Pwr Step.

The message “Mobile’s minimum power step > selected Power Cntl Step Size” will be displayed if a value is entered into the Power Cntl Step Size field that the Test Set determines is unacceptable

See Also

Protocol field description, on page 346 "MS Database" field on page 306

Screens Where Field is Present

CDMA TRANSMITTER POWER RANGE TEST

340

S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp

Page 340
Image 340