Agilent Technologies E8285A manual RF Gen Freq, RF Gen Volts, Sets the RF generator to 815 MHz

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 368
Image 368

Chapter 4, Description of Fields

Fields

RF Gen Freq

This field is displayed on the screens listed below when the RF Display field on the CONFIGURE screen is set to Freq.

This field sets the RF Generator’s (or CDMA generator’s) frequency.

GPIB Example

“RFG:FREQ 815MHZ“

sets the RF generator to 815 MHz.

See Also

RF Display field description, on page 367

Screens Where Field is Present

DUPLEX TEST

SPECTRUM ANALYZER (When the Controls field is set to RF Gen)

RX TEST

RF GENERATOR

CDMA CALL CONTROL

CDMA REVERSE CHANNEL SPECTRUM (When the Controls field is set to CDMA Gen)

RF Gen Volts

This field specifies whether you want RF voltages expressed as the voltage across a 50 ohm load, or the open circuit voltage (emf).

Operating Considerations

This setting affects the RF Generator’s and the Tracking Generator’s amplitudes.

Screens Where Field is Present

CONFIGURE

368

S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp

Page 368
Image 368
Agilent Technologies E8285A manual RF Gen Freq, RF Gen Volts, Sets the RF generator to 815 MHz