Chapter 4, Description of Fields

Fields

Closed Loop Pwr Cntl

This field sets the power control mode.

Open Loop mode causes the Test Set to send an alternating sequence of up/down power-control bits to the mobile station during periods when no other power-control operation is being performed. For example, if the Execute field is selected while the Test Set is in Open Loop mode, the mobile station will adjust its power according to the sequence of power control bits entered in the Change and Steps field. As soon as the number of power control bits entered in the Steps field have been sent, an alternating sequence of up/down power-control bits resumes, attempting to hold the mobile station’s power level at the value the mobile station was driven to by the Change and Steps fields.

NOTE:

Make sure that the Test Set power level is not reduced to a level where the mobile station

 

receives bad frames when using Open Loop power control mode. If the mobile station receives

 

a bad frame, it will assume the power control bit was a down bit and reduce its power by 1 dB.

 

When the power control mode is Open Loop, the alternating sequence of power control bits

 

will attempt to hold the mobile station’s power level at the reduced power level. As more bad

 

frames are received, the mobile station can continue to lower its power level until the call is

 

eventually dropped. See "Drop Timer" on page 238.

 

 

 

Off disables the power control subchannel.

Closed Loop sends power control bits to the mobile station in an attempt to drive the mobile station’s power level to a set-point based on the power transmitted from the Test Set and the open loop estimate.

Always Up forces the power control subchannel to send continuous “up” power control bits (0’s).

Always Down forces the power control subchannel to send continuous “down” bits. (1’s).

GPIB Example

“DISP CTXR;CDMA:PCONtrol:MODE ‘Open Loop’”

selects Open Loop power control mode.

Operating Considerations for Closed Loop Pwr Cntl

The power control operations performed when the Execute field is selected, such as “n up” or “ramp” take precedence over the power control bit sequence expected when a mode, such as “always up” is selected. For example, if the choice in the Change field is “n down” and the power control mode is “Always Up”, the number of “down” power control bits entered in the Steps field will be sent when the Execute field is selected. The Test Set will then resume sending continuous “up” power control bits.